Transvenous Pacing

Full Transvenous Pacemaker Setup:
1. 6F Cordis
2. Swan Ganz Pacing Catheter
3. Nonsterile Connecting Cable
4. Pacing Generator

Indications

  • Symptomatic sinus bradycardia (after atropine, +/- dopamine/epinephrine, and transcutaneous pacing have failed)
  • Mobitz type II second degree AV block
  • Complete heart block
  • Beta blocker or calcium channel blocker overdose

Preparation 

  • Patient positioning: supine/Trendelenburg
  • 6 French Cordis (“percutaneous sheath introducer kit”)
  • Swan Ganz/Bipolar pacing catheter
  • Pacer generator (“temporary pacemaker”)
  • Nonsterile connecting cable (within pacer generator case in inside sleeve)
  • Ultrasound + sterile probe cover

Supplies and room 9 location

Room 9 bay 1, bottom shelf on right
Swan Ganz/Bipolar pacing catheter
Pacer generator (Nonsterile connecting cable in the inside sleeve)
6 French Cordis (“percutaneous sheath introducer kit”)


Pacer generator—initial settings 

  • Turn on
  • Rate—80 bpm, rate at which patient will be transvenously paced, at least 20 bpm over the intrinsic rate
  • Output—20 mA, electrical output of pacer with every paced beat, decrease until patient has both mechanical (palpating patient’s pulse) and electrical capture
  • Sensitivity—3 mV, pacer’s ability to sense intrinsic rate (the lower the sensitivity, the more the pacer will detect intrinsic rate; for example, at 3 mV, the pacer will only detect impulses generated from the heart that are greater than 3 mV)
    • Oversensing- Sensitivity is set too LOW so electrical signals are inappropriately recognized as cardiac activity and pacing is inhibited
    • Undersensing- Sensitivity is set too HIGH so pacemaker ignores native cardiac activity

Location

  • Right internal jugular—preferred
  • Left subclavian—use as second option, preferred to leave site available for possible permanent pacer

Steps to placing transvenous pacemaker

  • Place cordis
  • Set up nonsterile connecting cable (helpful to have assistant connect cable to pacer generator as these are not sterile)
  • Ensure proper balloon inflation on Swan Ganz
  • Position sterile sleeve over pacer wire and ensure correct orientation
  • Insert pacer wire into cords and advance to 20cm (indicated by first two black lines)
  • Insert Swan-Ganz + and – pins into nonsterile connecting cable (proximal to positive, distal to negative)
  • Have assistant turn on pacer generator with the above settings (remember it is non sterile)
  • Inflate balloon and lock purple stopcock (stopcock is on Swan, catheter see below image)
  • Advance pacer wire to ~30-35cm (three black lines on pacer wire) while watching monitor for capture (STEMI pattern)
  • Troubleshooting: 
    • if wire coils in RA, pull wire back, twist 180˚ towards patient’s right and re-advance
    • if wire fails to capture, can adjust pacer generator settings, consider increasing output and decreasing sensitivity
  • Verify capture by either palpating pulse or by pulse ox waveform
  • Decrease output until there is no longer capture, then titrate up to the lowest effective output
  • Deflate balloon and turn stopcock off
  • Expand sterile sleeve
  • Suture cordis and place sterile dressing
  • Secure pacer wire to patient with tape (can dislodge easily)
Stopcock that comes attached to Swan-Ganz pacing catheter

Complications

  • Misplacement—can verify with ultrasound or chest x-ray
  • Ventricular perforation
  • Dysrhythmias
  • Pneumothorax

Leave a Reply

Your email address will not be published. Required fields are marked *