Indications
- Symptomatic sinus bradycardia (after atropine, +/- dopamine/epinephrine, and transcutaneous pacing have failed)
- Mobitz type II second degree AV block
- Complete heart block
- Beta blocker or calcium channel blocker overdose
Preparation
- Patient positioning: supine/Trendelenburg
- 6 French Cordis (“percutaneous sheath introducer kit”)
- Swan Ganz/Bipolar pacing catheter
- Pacer generator (“temporary pacemaker”)
- Nonsterile connecting cable (within pacer generator case in inside sleeve)
- Ultrasound + sterile probe cover
Supplies and room 9 location
Pacer generator—initial settings
- Turn on
- Rate—80 bpm, rate at which patient will be transvenously paced, at least 20 bpm over the intrinsic rate
- Output—20 mA, electrical output of pacer with every paced beat, decrease until patient has both mechanical (palpating patient’s pulse) and electrical capture
- Sensitivity—3 mV, pacer’s ability to sense intrinsic rate (the lower the sensitivity, the more the pacer will detect intrinsic rate; for example, at 3 mV, the pacer will only detect impulses generated from the heart that are greater than 3 mV)
- Oversensing- Sensitivity is set too LOW so electrical signals are inappropriately recognized as cardiac activity and pacing is inhibited
- Undersensing- Sensitivity is set too HIGH so pacemaker ignores native cardiac activity
Location
- Right internal jugular—preferred
- Left subclavian—use as second option, preferred to leave site available for possible permanent pacer
Steps to placing transvenous pacemaker
- Place cordis
- Set up nonsterile connecting cable (helpful to have assistant connect cable to pacer generator as these are not sterile)
- Ensure proper balloon inflation on Swan Ganz
- Position sterile sleeve over pacer wire and ensure correct orientation
- Insert pacer wire into cords and advance to 20cm (indicated by first two black lines)
- Insert Swan-Ganz + and – pins into nonsterile connecting cable (proximal to positive, distal to negative)
- Have assistant turn on pacer generator with the above settings (remember it is non sterile)
- Inflate balloon and lock purple stopcock (stopcock is on Swan, catheter see below image)
- Advance pacer wire to ~30-35cm (three black lines on pacer wire) while watching monitor for capture (STEMI pattern)
- Troubleshooting:
- if wire coils in RA, pull wire back, twist 180˚ towards patient’s right and re-advance
- if wire fails to capture, can adjust pacer generator settings, consider increasing output and decreasing sensitivity
- Verify capture by either palpating pulse or by pulse ox waveform
- Decrease output until there is no longer capture, then titrate up to the lowest effective output
- Deflate balloon and turn stopcock off
- Expand sterile sleeve
- Suture cordis and place sterile dressing
- Secure pacer wire to patient with tape (can dislodge easily)
Complications
- Misplacement—can verify with ultrasound or chest x-ray
- Ventricular perforation
- Dysrhythmias
- Pneumothorax